[Frontiers in Bioscience E4, 1731-1742, January 1, 2012] |
|
|
Evaluation of unintended electrical stimulation from MR gradient fields Howard I. Bassen1, Leonardo M. Angelone1
1 TABLE OF CONTENTS
1. ABSTRACT Exposure of patients with active implants (e.g. cardiac pacemakers and neurostimulators) to magnetic gradient fields (kHz range) during magnetic resonance imaging presents safety issues, such as unintended stimulation. Magnetically induced electric fields generate currents along the implant's lead, especially high at the distal tip. Experimental evaluation of the induced electric field was previously conducted. This study aimed to perform the same evaluation by means of computational methods, using two commercially available software packages (SemcadX and COMSOL Multiphysics). Electric field values were analyzed 1-3 mm from the distal tip. The effect of the two-electrode experimental probe was evaluated. The results were compared with previously published experimental data with reasonable agreement at locations more than 2-3 mm from the distal tip of the lead. The results were affected by the computational mesh size, with up to one order of magnitude difference for SEMCAD (resolution of 0.1 mm) compared to COMSOL (resolution of 0.5 mm). The results were also affected by the dimensions of the two-electrode probe, suggesting careful selection of the probe dimensions during experimental studies. 2. INTRODUCTION Magnetic resonance imaging (MRI) of patients with active implants, such as cardiac pacemakers and neurological stimulators, presents safety issues that need to be carefully addressed (1). One of these safety issues is due to electric potential induced along the implants by the low-frequency - kHz range - MRI gradient fields. These potentials generate currents along the lead and an enhancement of the electric field near the distal tip of the lead (2). The issue is also present with other common devices that emit magnetic fields in the same frequency range as gradient MRI, including metal detectors and anti-theft systems. Over 100 adverse events related to unintended stimulation were reported to FDA and as a result FDA issued official alerts to cardiologists, neurologists and other clinicians (3). Methods for evaluating the electric field induced by strong (i.e. ~ 0.1 Tesla) low-frequency magnetic fields have been studied in detail in experimental studies. Our group has recently published the measured data on the magnetically induced electric field near the distal tip of implanted leads (4). The study evaluated the electric field induced by a 1 kHz homogeneous magnetic field at the distal tip of a lead in a saline tank, simulating a unipolar pacemaker lead in cardiac tissue (Figure 1). Computational modeling of the experimental measurement system - properly validated with measured data - is highly desirable since it allows extrapolating the results to many configurations that would otherwise be tedious or too complex to test experimentally (5, 6). The objective of this study was to implement a numerical model similar to the experimental system, compute electric field and currents induced along the lead by the low-frequency magnetic field using two commercially available software packages, and compare the results with the measured values. A realistic numerical modeling of an active implant lead in saline or in a human body is challenging for several reasons. The geometrical characteristics of the lead, namely a length greater than 0.5 m and a diameter of the wires inside the lead less than 1 mm, require a variable grid size. The steep change of electric field - which at few millimeters from the distal tip decays to insignificant levels - requires a submillimetric resolution to model the tip and the nearby conductive medium (i.e., saline). Finally, the modeling of the thin insulation layer requires a large number of computational cells. Recent improvements in electromagnetic software and reduced cost of RAM memory allowed to overcome these limitations and to generate reasonably accurate models. 3. METHODS 3.1. SEMCAD simulations A first set of simulations was carried out using SemcadX version 14 (SPEAG, Zurich, Switzerland) with a new Low Frequency Quasi Static Magnetic Solver. This solver uses a finite element method based on the Biot-Savart equation and an adaptive mesh gridding engine. This allowed analyzing very small objects with a submillimetric resolution for selected parts of a thin lead, while encompassing the large diameter loop in a saline tank. Computations were performed with a personal computer with 6 GB RAM and Windows 7 operating system. Numerical model of saline solution with lead The leads were modeled by an insulated wire with a 0.54 m length immersed in a saline tank. The tank had a diameter of 0.24 m and a height of 0.23 m. The electrical properties of the saline were set to < Magnetic field source The magnetic field source was modeled by a pair of single turn coils in the form of a modified Helmholtz coil (8) driven by a 1 kHz sinusoidal current of 34.7 A, simulating the output of the high-current amplifier used in the experimental measurements (4). This produced a homogeneous magnetic field in the saline with a constant magnitude of < Electromagnetic simulations The gridded model of the system without the electric field probe is shown in Figure 2. Areas that did not require fine detail (e.g. the saline) were meshed with a large grid (i.e., tens of mm), while the insulation around the lead and the lead wire itself were meshed with a variable sized grid with the finest resolution up to 10-4 m. This allowed to drastically reducing the number of solid rectangular voxels and cells that needed to be analyzed compared to a fixed mesh approach. Figure 2 illustrates the high resolution of the voxels generated for the insulation and wire plus the electric field probe electrodes. No breakage in the continuity of the lead wire existed, and the insulation had no gaps that could expose the wire to the surrounding saline. A range of 6 to 18 Megacells was used in the simulations to identify the most reliable data, with corresponding simulation times between 30 and 100 minutes, respectively. 3.2. COMSOL simulations A second set of electromagnetic simulations was performed using a commercially available Finite Element Method solver (COMSOL Multiphysics 3.5, Comsol Inc., Burlington, MA, USA) and a 16-processor Xeon 2.4 GHz machine with 24 GB RAM and 64-bit Windows Server Standard operating system. Numerical model of saline solution with lead A saline solution was modeled by means of a conductive cylinder (diameter: 0.24 m, height: 0.1 m - see discussion on modeling the saline tank) (Figure 3). The electrical properties of the saline were set to < Magnetic field source A homogeneous 1 kHz sinusoidal magnetic flux density perpendicular to the base of the saline cylinder was assigned as boundary condition and set to a constant value of < Electromagnetic simulations The quasi-static electromagnetic solver of the COMSOL Multiphysics - with electric and induced current - was selected for the simulations. The following equations were implemented: < where < 3.3. Comparison to analytical solution and validation of model without lead In order to provide a simple validation of the model, the magnitude of the electric field < < where < 4. RESULTS The magnitude of the magnetic flux density < The RMS magnitude of electric field < 5. DISCUSSION 5.1. Modeling the magnetic field source in SEMCAD and COMSOL The magnetic flux density < 5.2. Modeling the distal tip, proximal tip, and stimulator pulse generator The distal tip of the lead in both SEMCAD and COMSOL models was bluntly cut flush with the insulation. This modeled a worst case scenario, since the previous experimental study (4) showed that a distal tip bluntly cut flush with the insulation produced a higher electric field in the adjacent saline when compared to a distal tip protruding past the insulation. The metallic case of the stimulator pulse generator was not included in the computational model and the proximal end of the wire had insulation removed for 5∙10-3 m, serving as the reference (return or ground) electrode (Figure 2 and Figure 3). The previous experimental study showed that this resulted in the same electric field strength adjacent to the distal tip compared to a connection to the case of an actively pulsing pacemaker stimulator (4). 5.3. Modeling the lead insulation The diameter of the wire was 6∙10-4 in both SEMCAD and COMSOL, modeling the same lead diameter used in the experimental study. The insulation of the lead modeled in SEMCAD was 4∙10-3 m, to avoid that - with a fixed spatial resolution of the modeled grid - parts of the lead were not directly in contact with the saline tank. To test the effect of the insulation, both 1.4∙10-3 m and 4∙10-3 m insulation models were tested in COMSOL, as in the experimental study and in SEMCAD, respectively. The COMSOL results showed a 20% - 30% effect of these values of insulation on the overall electric field at a 2∙10-3 m distance from the distal tip (Table 1). 5.4. Effect of mesh size and reduced resolution on COMSOL simulations Because of RAM memory requirements, the minimum mesh size produced by COMSOL was approximately 5∙10-3 m near the distal tip. This was in contrast to the submillimetric resolution used in SEMCAD with less physical computer (RAM) memory. As a consequence of this limitation, the closest distance from the distal tip where the electric field could be accurately analyzed with COMSOL was 1 mm. 5.5. Computation of electric field at the tip of the simulated lead and comparison with experimental results The maximum resulting induced electric field < 5.6. Computational uncertainty There are several sources of uncertainties/errors that occur when measuring highly localized electric fields with a probe of finite dimensions, including spatial averaging, field perturbation, probe positioning, and electromagnetic pickup. The previous experimental study (4) pointed out measurement uncertainties of electric fields when the region of interest was extremely close to the tip of a simulated or actual stimulator lead. 5.6.1. Spatial averaging One source of uncertainty is related to the changes of electric fields in space. The electric field magnetically induced at the tip of a lead has large spatial gradients, with significant changes in magnitude over a single millimeter. Spatial averaging of the electric field occurs over the region occupied by the probe tips (9∙10-4 m wide x 3.5∙10-3 m deep in (4).) This averaging produces a reduction in the measured value of the electric field compared to the true field point value. The uncertainty is greatest when the distance between the probe and the lead is the same or less than the probe tip size. 5.6.2. Field perturbation Another type of measurement uncertainty is due to perturbation of the electric field by metallic probe electrodes. This uncertainty is greatest when the distance between the probe and the lead is the same or less than the probe tip size. This perturbation makes the measured value of the electric field lower than the true value. This uncertainty diminishes with increasing separation between the probe and the tip of the lead. 5.6.3. Uncertainties due to probe positioning A third measurement uncertainty is due to the error and repeatability associated with determining the true position of the probe tip with respect to an implant lead. For the mechanical system used in (4), the positioning error and repeatability was about 10-4 m. The measurement uncertainty due to this factor diminished as the separation between the probe and the implant lead increased. As shown in Figure 6, a change of 10-4 m in position generates a change of the computed electric field value of about 20%. 5.7. Effect of the presence of a physical measuring probe As shown in (4), the probe used to measure the electric field introduces certain measurement uncertainties. In order to understand the effects of field perturbation by the probe, a computational model that included a two-electrode probe was generated with SEMCAD. The probe closest electrode was placed at various distances from the distal tip of the simulated implant lead wire. The probe was modeled in SEMCAD as two short, parallel wires (or electrodes) each having a diameter of 2∙10-4 m and spaced 5∙10-4 m apart. The electrical properties of the electrodes were chosen to model a perfect electrical conductor. The best matches - based on the limits of the Low Frequency Quasi Static Magnetic Solver in SemcadX - were conductivity < 6. CONCLUSIONS There is a significant need to evaluate the safety of implanted medical devices utilizing long, thin, insulated stimulation leads or conductive wires exposed to kHz-range magnetic fields, to avoid unintended cardiac or neural stimulation. This is often accomplished by evaluating the electric field with invasive connections of wires made inside the implanted device or its lead connectors. However, these invasive modifications can only be made by the device manufacturers. Even when invasive wires are attached, they modify the magnetic field coupling to the device under test, adding measurement uncertainties for unintended stimulation. This study of the electric field distribution induced by low-frequency sinusoidal magnetic fields used non-invasive computational methods, and compared the results to prior experimental measurements with a two-electrode probe. The electric field was computed at locations of few mm from the distal tip of a lead immersed in a saline tank. Two commercially available electromagnetic solvers were used, namely SemcadX and COMSOL Multiphysics, both based on the finite element method. Reasonable agreement in the electric field values obtained by the two computational models existed at a distance of two or more mm from the lead, with uncertainties due spatial averaging related to the numerical grid used in the simulations. The submillimetric resolution used in SEMCAD with less physical computer (RAM) memory allowed calculating the electric field at distances of up to 1 mm from the distal tip. In the experimental study, the probe outer dimensions were 9∙10-4 m; therefore the experimental data measured at distances closer than 1 mm was not accurate. The computational models agreed well (�10%) with the experimental results when the electric field measurement probes were included in the model and their effects on the local electric field were therefore accounted for. Measurements with smaller probes may provide an accurate value of magnetically induced electric fields at distances closer than 1 mm from the distal tip. 7. ACKNOWLEDGMENTS The mention of commercial products, their sources, or their use in connection with material reported herein is not to be construed as either an actual or implied endorsement of such products by the Department of Health and Human Services. The authors would like to thank Seth Seidman for the useful comments and insights and Isaac Chang for the support with the COMSOL simulations. 8. REFERENCES 1. E. T. Martin, J. A. Coman, F. G. Shellock, C. C. Pulling, R. Fair, and K. Jenkins, Magnetic resonance imaging and cardiac pacemaker safety at 1.5-Tesla. J Amer Coll Cardiol, 43, 1315-1324 (2004).
2. A.W. Guy, Calculation of Field Enhancement due to external leads and implants in contact with Tissues. In Appendix G, NCRP Report 67: Radiofrequency Electromagnetic Fields: Properties, Quantities and Units, Biophysical Interaction, and Measurements, 108-111 (1981).
3. D. B. Burlington. Important Information on Anti-Theft and Metal Detector Systems and Pacemakers, ICDs, and Spinal Cord Stimulators. Center for Devices and Radiological Health, Food and Drug Administration, September 28, 1998
4. HI Bassen and GG Mendoza, In-vitro mapping of E-fields induced near pacemaker leads by simulated MR gradient fields. BioMedical Engineering OnLine 8:39 (2009)
6. CM Collins, Numerical field calculations considering the human subject for engineering and safety assurance in MRI. NMR Biomed 22(9):919-26 (2009).
7. A. Stogryn, Equations for calculating the dielectric constant of saline water. IEEE Trans on MicrowaveTheory and Techniques, August, pg 733-736 (1971).
8. V. Buzduga, DM Witters, JP Casamento, W Kainz, Testing the immunity of active implantable medical devices to CW magnetic fields up to 1 MHz by an immersion method. IEEE Trans Biomed Eng 54 (9), pp. 1679-1686, (2007). Key Words: Pacemaker, Neurostimulator, Electric Field Enhancement, Spatial Resolution, Probe, kHz, Electromagnetism Send correspondence to: Howard I. Bassen, 10903 New Hampshire Avenue, WO62 Rm. 1112, Silver Spring, MD 20993-0002. Tel: 301-796-2595, Fax: 301-796-9927, E-mail:howard.bassen@fda.hhs.gov |